In order to match multiple cuvettes follow the next
steps:
Fill some cuvettes with high quality water
(<0.1 NTU) up to the line.
Clean and oil the cuvettes as described before.
Turn the instrument ON.
Insert the first cuvette into the instrument and
press READ u. Record the reading.
Record the position of the cuvette and the
displayed reading.
Mark this position on the thicker white band on
the top of the cuvette with a water resistant
pencil.
Insert the second cuvette into the instrument
and take a reading.
MATCHING MULTIPLE CUVETTES
Precise measurements require the use of a single cuvette. If it is not possible, the cuvette
selection and matching must be performed before taking measurements.