19
Notes: When the instrument is not calibrated or calibration is cleared
(default values loaded) there is no “expired calibration”,
and the display always shows the “CAL” “INTV” tags blinking.
When an abnormal condition in the RTC is detected, the
instrument forces the “expired calibration” status.
pH CALIBRATION DATA
Calibration data is stored automatically after a successful calibration.
To view the pH calibration data, press GLP while in measurement mode.
The instruments will display the time (hh:mm) of the last calibration.
Use the ARROW keys to scroll through the calibration data:
The date (yyyy.mm.dd).
The pH calibration offset.
The pH calibration slope (the GLP slope is the average of the calibration
slopes; the percentage is referred to the ideal value of 59.16 mV/pH).
The pH calibration buffers in calibrating order and with the selected
resolutions at calibration moment.
The first pH calibration buffer:
The second pH calibration buffer: