Hanna-instruments HI 4101 Manuale Utente Pagina 15

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 40
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 14
15
Typical calibration curve for HI 4101 Ammonia ISE
Standards should bracket the measurement range of
interest and differ from each other by a factor of 10 in
the linear regions.
Standards and solutions should be at the same tem-
perature. 2 mL of ISA is added to each 100 mL of
sample and standard. A color change should occur to
indicate the ISA has been added.
Protect these solutions from loss of dissolved gas by
covering and using promptly.
3) Follow section VII; General Guidelines to optimize test
set-up.
4) During calibration it is best to start with lower concen-
tration samples first. Wait for a stable reading before
reading/recording values.
5) To prevent carry over and contamination of samples,
rinse sensors with deionized water and blot dry be-
tween samples.
6) Between measurements suspend sensor tip in a small
sample of NH
3
Conditioning solution;
HI 4001-45. Rinse body with deionized water and
blot dry before placing in next sample.
Typical Linearity for HI 4101 Electrode
-300
-250
-200
-150
-100
-50
0
50
0123456
-Log of [M
]
m
V
Vedere la pagina 14
1 2 ... 10 11 12 13 14 15 16 17 18 19 20 ... 39 40

Commenti su questo manuale

Nessun commento