Reinsert the cuvette into the instrument.
Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait
for 6 minutes and press Read. When the timer ends the meter will perform the reading. The instrument
displays concentration in mg/L of nitrite.
Press or to access the second level functions.
Press the Chem Frm key to convert the result in mg/L of nitrogen-nitrite (NO
2
¯-N) and sodium nitrite
(NaNO
2
).
Press or to return to the measurement screen.
INTERFERENCES
Interference may be caused by the following ions:
ferrous, ferric, cupric, mercurous, silver, antimonious, bismuth, auric, lead, metavanadate and chloroplatinate.
Strongly reducing and oxidizing reagents.
High levels of nitrate (above 100 mg/L) could yield falsely high readings due to a minute amount of
reduction to nitrite that could occur at these levels.
Nitrite LR