Ammonia HR
Remove the cuvette.
Add 4 drops of HI 93733A-0 Nessler Reagent. Replace
the cap and mix the solution.
Reinsert the cuvette into the instrument.
Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait
for 3 minutes and 30 seconds and press Read. When the timer ends the meter will perform the
reading. The instrument displays the results in mg/L of ammonium nitrogen (NH
3
-N).
Press or to access the second level functions.
Press the Chem Frm key to convert the result in mg/L of ammonia (NH
3
) and ammonium (NH
4
+
).
Press or to return to the measurement screen.
INTERFERENCES
Interference may be caused by:
Acetone, alcohols, aldehydes, glycine, hardness above 1 g/L, iron, organic chloramines, sulfide, various
aliphatic and aromatic amines.