5-24
METHODS
5.5.14.9 Flow Rate
The flow rate for the dosing system can be set by the user in an interval of 0.1 to two times
the burette volume: 0.1 to 10 mL/min for a 5 mL burette
Note: The titrator will automatically detect the burette size and display the correct high limit
volume.
The flow rate is set for all burette operations.
5.5.15 Termination Parameters
This screen allows the user to set the control parameters related to the end of the titration.