5-20
METHODS
5.5.11.1 Maximum Duration
Specify the maximum time a titration is allowed to run. Once this point is reached, the
titration will be terminated even if the end point is not reached.
The time can be set from 10 to 3600 seconds.
5.5.11 Termination Parameters
This screen allows the user to set the control parameters related to the end of the titration.
Vedere la pagina 77