Hanna-instruments HI 93414 Manuale Utente Pagina 17

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 52
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 16
17
In order to match multiple cuvettes follow the next
steps:
Fill some cuvettes with high quality water
(<0.1NTU) up to the 10 mL mark.
Clean and oil the cuvettes as described before.
Turn the instrument ON.
Insert the first cuvette into the instrument and
press READ/TIMER u. Record the reading.
Record the position of the cuvette and the
displayed reading.
Mark this position on the thicker white band on
the top of the cuvette with a water resistant
pencil.
Insert the second cuvette into the instrument
and take a reading.
MATCHING MULTIPLE CUVETTES
Precise measurements require the use of a single cuvette. If it is not possible, cuvette selection
and matching must be performed before taking measurements.
Vedere la pagina 16
1 2 ... 12 13 14 15 16 17 18 19 20 21 22 ... 51 52

Commenti su questo manuale

Nessun commento