Hanna-instruments HI 95715 Manuale Utente Pagina 14

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 20
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 13
14
Replace the cuvet into the holder and ensure
that the notch on the cap is positioned securely
into the groove.
Hold READ/TIMER for three seconds. The
display will show the countdown prior to
measurement.
Alternatively, wait for 3 minutes and 30 seconds
and just press READ/TIMER.
In both cases “SIP” will blink during
measurement.
The instrument directly displays concentration
in mg/L of ammonia nitrogen (NH
3
-N) on the
Liquid Crystal Display.
To convert the reading to mg/L of ammonia
(NH
3
), multiply by a factor of 1.216.
or
INTERFERENCES
Organic compounds like: chloramines, various aliphatic and aromatic amines, glycine,
(positive error).
To eliminate these interferences distillation is required.
Organic compounds like: aldehydes, alcohols (e.g. ethanol) or acetone (negative error).
To eliminate these interferences distillation is required.
Sulfide (S
2-
): may cause turbidity.
Hardness above 1 g/L as Calcium Carbonate (CaCO
3
).
Vedere la pagina 13
1 2 ... 9 10 11 12 13 14 15 16 17 18 19 20

Commenti su questo manuale

Nessun commento