Hanna-instruments HI 96704 Manuale Utente Pagina 16

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 24
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 15
16
Press CAL CHECK and the lamp, cuvette and
detector icons together with “CAL CHECK”
will appear on the display, depending on the
measurement phase.
At the end of the measurement the display
will show the validation standard value.
The reading should be within specifications as reported in the CAL CHECK
TM
Standard
Certificate. If the value is found out of the specifications, please check that the cuvettes are free
of fingerprints, oil or dirt and repeat validation. If results are still found out of specifications,
then recalibrate the instrument.
Vedere la pagina 15
1 2 ... 11 12 13 14 15 16 17 18 19 20 21 22 23 24

Commenti su questo manuale

Nessun commento