Hanna Instruments Thermometer HI 7034 Manuale Utente Pagina 10

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 20
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 9
10
Press the COND/TEMP key to display conduc-
tivity.
The display should read the conductivity of the
calibrating solution at the reference tempera-
ture.
E.g. using HI7030/HI8030 with a so-
lution at 18°C, the reading will be
"12.88 mS".
PROCEDURE FOR HI8820:
Pour a small quantity of the conductivity solu-
tion into a plastic beaker.
Immerse the conductivity probe in the solution sub-
merging the holes of the sleeve (0.5 cm below).
Make sure that no air bubbles are trapped inside
the glass sleeve. Wait for 2 or 3 minutes for thermal
equilibrium.
Switch the instrument on by pressing the ON/OFF
key.
Select the appropriate conductivity range
e.g. "19.99 mS" for HI7030/HI8030,
"99.9 mS" for HI7034/HI8034,
"1999µS" for HI7031/HI8031,
"199.9 µS" for HI7033/HI8033.
Set the temperature coefficient knob to
2%°C.
HI 7030
0
0.5
1
1.5
2
2.5
TEMPERATURE
COEFFICIENT
Vedere la pagina 9
1 2 ... 5 6 7 8 9 10 11 12 13 14 15 ... 19 20

Commenti su questo manuale

Nessun commento