CALIBRATION DATA CALIBRATION DATA
CALIBRATION DATA CALIBRATION DATA
CALIBRATION DATA
Relative mV calibration data is stored automatically after a successful
calibration.
To view the Relative mV calibration data, press GLP when the instrument
is in Relative mV measurement mode.
The instrument will display the time (hh:mm) of the last calibration.
Use the ARROW keys to scroll through the calibration data:
The date (yyyy:mm:dd).
The Relative mV calibration offset.
• The instrument ID.
INTV
CAL
TIME
CAL
- or the number of days until the calibration alarm will be displayed.
- or if expired (7 days ago).
• The instrument ID.