50
GLDxxxff: Requests the records of the “xxx” lot number. The records are
sent in frames of 10 records; “ff” is the frame number (01 first
frame). (Example: Lot 13 has 53 records. The records will be
sent in 6 frame, 5 with10 records and 1 with 3 records.)
Command Parameters:
xxx - Lot number (eq: for lot number 1 xxx = 001)
ff - requested frame number - first frame is labeled 01
Errors: “Err3” log on demand empty.
“Err4” requested set parameter is not available.
“Err5” command argument is wrong.
“Err6” requested range not available.
“Err7” meter in log mode.
“Err8” is sent if instrument is not in measurement mode.
“NAK” (21) character is sent when the instrument receives
an unknown or a corrupted command.
Note: P - request for pH range.
M - request for mV range.
E - request for EC range.
T - request for TDS range.
N - request for NaCl range.