Hanna-instruments HI 3220 Manuale Utente Pagina 26

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 25
26
FIVE-POINT CALIBRATION (HI 3222 only)
Use part of this procedure for 2, 3 or 4 point calibration.
Press ESC.
Immerse the ISE electrode approximately 2 cm (¾”) into
the less concentrated standard solution and stir gently.
Press CAL. The primary LCD will displays the ion concentration in the
selected unit or “---” if not calibrated and first standard value.
If necessary, press the ARROW keys to select a different standard value.
The “ ” tag will blink on the LCD until the reading is stable.
When the reading is stable and close to the selected standard, the Confirm
function key is displayed.
Press Confirm to confirm calibration.
The calibrated value and the second expected standard value will be
displayed.
After the first calibration point is confirmed, immerse the ISE electrode
approximately 2 cm (¾”) into the second calibration solution.
If necessary, press the ARROW keys to select a different standard value.
The “ ” tag will blink on the LCD until the reading is stable.
When the reading is stable and close to the selected standard, the
Confirm function key is displayed.
Press Confirm to confirm calibration.
The calibrated value and the third expected standard value will be
displayed.
2 cm
Vedere la pagina 25
1 2 ... 21 22 23 24 25 26 27 28 29 30 31 ... 67 68

Commenti su questo manuale

Nessun commento