the average of slopes, with sign and decimal point (7 chars)
the calibration time, yymmddhhmmss (12 chars)
buffers information (for each buffer)
type (1 char): 0 - standard, 1 - custom
status (1 char): N (new) - calibrated in last calibration;
O (old) - from an old calibration.
warnings during calibration (2 chars): 00 - no
warning, 04 - Clean Electrode warning.
buffer value, with sign and decimal point and expo-
nent (11 chars).
calibration time, yymmddhhmmss (12 chars).
• electrode condition, with sign (3 chars). The “-01”
code means not calculated.
Rel mV calibration data (if available), which contains:
the calibration offset, with sign (7 chars)
the calibration time, yymmddhhmmss (12 chars).
ISE calibration data (if available) when x=1 (HI 3221)
or when x=2 (HI 3222), which contains:
the number of calibrated standards (1 char)
the ion charge, with sign (2 chars) (HI 3221)
the calibration slope, with sign and decimal point (7 chars)
the calibration time, yymmddhhmmss (12 chars)
standards information (for each standard)
type (1 char): 0 - always standard solution.
status (1 char):N (new) - calibrated in last calibration;
O (old) - from an old calibration.
warnings during calibration (2 chars): 00 - no warning.
standard value, with sign and decimal point and
exponent (11 chars).
calibration time, yymmddhhmmss (12 chars).
PARx Requests the setup parameters setting.
x=1 setup parameters for channel 1
x=2 setup parameters for channel 2 (HI 3222)
The answer string contains:
Backlight Value (1 ASCII char)
Contrast Value (2 ASCII char)
Instrument ID (4 chars)
Calibration Alarm time out for pH (2 chars) when x=1